Rocksolid Light

Welcome to Rocksolid Light

mail  files  register  newsreader  groups  login

Message-ID:  

What is now proved was once only imagin'd. -- William Blake


tech / sci.electronics.repair / Failure of a pair of Eppindorf centrifuges.

SubjectAuthor
* Failure of a pair of Eppindorf centrifuges.peter
+- Failure of a pair of Eppindorf centrifuges.Peter W.
+- Failure of a pair of Eppindorf centrifuges.Bob Engelhardt
`- Failure of a pair of Eppindorf centrifuges.whit3rd

1
Failure of a pair of Eppindorf centrifuges.

<ujm1f5$1h6c6$1@dont-email.me>

  copy mid

https://news.novabbs.org/tech/article-flat.php?id=4550&group=sci.electronics.repair#4550

  copy link   Newsgroups: sci.electronics.repair
Path: i2pn2.org!i2pn.org!eternal-september.org!feeder3.eternal-september.org!news.eternal-september.org!.POSTED!not-for-mail
From: peter@easthope.ca
Newsgroups: sci.electronics.repair
Subject: Failure of a pair of Eppindorf centrifuges.
Date: 22 Nov 2023 15:05:09 -0700
Organization: A noiseless patient Spider
Lines: 18
Message-ID: <ujm1f5$1h6c6$1@dont-email.me>
MIME-Version: 1.0
Content-Type: text/plain; charset=ISO-8859-1
Content-Transfer-Encoding: 8bit
Injection-Date: Wed, 22 Nov 2023 23:05:09 -0000 (UTC)
Injection-Info: dont-email.me; posting-host="e8fa8b42423b3f616c3f60e1030b82de";
logging-data="1612166"; mail-complaints-to="abuse@eternal-september.org"; posting-account="U2FsdGVkX19229TG1rOStBQ6xoP9uZhA"
Cancel-Lock: sha1:FlLk5T6MeXxCzk2UwAsqtPspU1o=
X-Mailer: Oberon Mail (ejz) on LinuxA2 Gen. 32-bit, rev.9799
 by: peter@easthope.ca - Wed, 22 Nov 2023 22:05 UTC

Hi,

The photo here is one of a pair of Eppindorf 5415D centrifuges.
http://easthope.ca/Eppindorf5415D.jpg

This morning both machines failed to operate. No response when the
main power switch is flipped on. Empty display. No power indicator
light. Fuses are OK.

The machines are in separate rooms. There is power at the receptacles
they connect to. No other equipment is failing.

Ideas?

Thanks, ... P.

Re: Failure of a pair of Eppindorf centrifuges.

<dc665acf-318b-4c53-8c20-a0f3e25f7b46n@googlegroups.com>

  copy mid

https://news.novabbs.org/tech/article-flat.php?id=4551&group=sci.electronics.repair#4551

  copy link   Newsgroups: sci.electronics.repair
X-Received: by 2002:ac8:5c0e:0:b0:417:fa13:5bbb with SMTP id i14-20020ac85c0e000000b00417fa135bbbmr158945qti.8.1700742818581;
Thu, 23 Nov 2023 04:33:38 -0800 (PST)
X-Received: by 2002:ad4:4304:0:b0:679:cf82:e6fd with SMTP id
c4-20020ad44304000000b00679cf82e6fdmr139993qvs.0.1700742818338; Thu, 23 Nov
2023 04:33:38 -0800 (PST)
Path: i2pn2.org!i2pn.org!news.neodome.net!weretis.net!feeder6.news.weretis.net!usenet.blueworldhosting.com!diablo1.usenet.blueworldhosting.com!peer02.iad!feed-me.highwinds-media.com!news.highwinds-media.com!news-out.google.com!nntp.google.com!postnews.google.com!google-groups.googlegroups.com!not-for-mail
Newsgroups: sci.electronics.repair
Date: Thu, 23 Nov 2023 04:33:37 -0800 (PST)
In-Reply-To: <ujm1f5$1h6c6$1@dont-email.me>
Injection-Info: google-groups.googlegroups.com; posting-host=2600:4040:7cd2:d700:1429:e68d:74c8:c8e3;
posting-account=tQE80AoAAACYibCqNWmwmvQ2q1XctZSS
NNTP-Posting-Host: 2600:4040:7cd2:d700:1429:e68d:74c8:c8e3
References: <ujm1f5$1h6c6$1@dont-email.me>
User-Agent: G2/1.0
MIME-Version: 1.0
Message-ID: <dc665acf-318b-4c53-8c20-a0f3e25f7b46n@googlegroups.com>
Subject: Re: Failure of a pair of Eppindorf centrifuges.
From: peterwieck33@gmail.com (Peter W.)
Injection-Date: Thu, 23 Nov 2023 12:33:38 +0000
Content-Type: text/plain; charset="UTF-8"
X-Received-Bytes: 1985
 by: Peter W. - Thu, 23 Nov 2023 12:33 UTC

Same age? Close serial numbers?

As to power - did you plug in another piece of equipment, or just check with some sort of tester? If the neutral is out, you may have "power", but not usable power.

Suggestions in no particular order:

a) Find the main panel that feeds those circuits and the individual breaker(s). Cycle them.
b) Move the machines to another known-good receptacle.
c) Check the power cords. If they are not hard-wired, try removing/plugging back in.
d) Check the lid interlock(s). If they are sticking, the machines will not start. In the case of some of the machines here, they will not even light up if the interlock is not either fully open or fully engaged.
e) Replace the fuses anyway. They get tired and may not show as blown, but may still be open.

Good Luck!

Peter Wieck
Melrose Park, PA

Re: Failure of a pair of Eppindorf centrifuges.

<63J7N.89982$AqO5.51427@fx11.iad>

  copy mid

https://news.novabbs.org/tech/article-flat.php?id=4552&group=sci.electronics.repair#4552

  copy link   Newsgroups: sci.electronics.repair
Path: i2pn2.org!i2pn.org!weretis.net!feeder8.news.weretis.net!feeder1.feed.usenet.farm!feed.usenet.farm!peer01.ams4!peer.am4.highwinds-media.com!peer03.iad!feed-me.highwinds-media.com!news.highwinds-media.com!fx11.iad.POSTED!not-for-mail
Subject: Re: Failure of a pair of Eppindorf centrifuges.
Newsgroups: sci.electronics.repair
References: <ujm1f5$1h6c6$1@dont-email.me>
From: BobEngelhardt@comcast.net (Bob Engelhardt)
User-Agent: Mozilla/5.0 (Windows NT 10.0; WOW64; rv:52.0) Gecko/20100101
Thunderbird/52.6.0
MIME-Version: 1.0
In-Reply-To: <ujm1f5$1h6c6$1@dont-email.me>
Content-Type: text/plain; charset=utf-8; format=flowed
Content-Language: en-US
Content-Transfer-Encoding: 7bit
Lines: 9
Message-ID: <63J7N.89982$AqO5.51427@fx11.iad>
X-Complaints-To: https://www.astraweb.com/aup
NNTP-Posting-Date: Thu, 23 Nov 2023 14:10:10 UTC
Date: Thu, 23 Nov 2023 09:10:10 -0500
X-Received-Bytes: 1257
 by: Bob Engelhardt - Thu, 23 Nov 2023 14:10 UTC

On 11/22/2023 5:05 PM, peter@easthope.ca wrote:
>[...]
> The machines are in separate rooms. There is power at the receptacles
> they connect to. No other equipment is failing.

"Separate rooms"!! That is truly confounding. Ordinarily, the 2 rooms
situation would imply 2 power circuits (breakers). But with identical,
spontaneous failures of both machines, they are _most likely_ on the
same circuit. So the breaker is tripped, or defective.

Re: Failure of a pair of Eppindorf centrifuges.

<f50262de-f355-4b42-a22b-6fda17c4e4e5n@googlegroups.com>

  copy mid

https://news.novabbs.org/tech/article-flat.php?id=4553&group=sci.electronics.repair#4553

  copy link   Newsgroups: sci.electronics.repair
X-Received: by 2002:ac8:47c3:0:b0:41e:311a:77ac with SMTP id d3-20020ac847c3000000b0041e311a77acmr1758qtr.11.1700762218258;
Thu, 23 Nov 2023 09:56:58 -0800 (PST)
X-Received: by 2002:a05:6a00:3927:b0:690:2fa3:9769 with SMTP id
fh39-20020a056a00392700b006902fa39769mr51180pfb.5.1700762217942; Thu, 23 Nov
2023 09:56:57 -0800 (PST)
Path: i2pn2.org!i2pn.org!weretis.net!feeder8.news.weretis.net!feeder1.feed.usenet.farm!feed.usenet.farm!peer03.ams4!peer.am4.highwinds-media.com!peer01.iad!feed-me.highwinds-media.com!news.highwinds-media.com!news-out.google.com!nntp.google.com!postnews.google.com!google-groups.googlegroups.com!not-for-mail
Newsgroups: sci.electronics.repair
Date: Thu, 23 Nov 2023 09:56:57 -0800 (PST)
In-Reply-To: <ujm1f5$1h6c6$1@dont-email.me>
Injection-Info: google-groups.googlegroups.com; posting-host=209.221.140.126; posting-account=vKQm_QoAAADOaDCYsqOFDAW8NJ8sFHoE
NNTP-Posting-Host: 209.221.140.126
References: <ujm1f5$1h6c6$1@dont-email.me>
User-Agent: G2/1.0
MIME-Version: 1.0
Message-ID: <f50262de-f355-4b42-a22b-6fda17c4e4e5n@googlegroups.com>
Subject: Re: Failure of a pair of Eppindorf centrifuges.
From: whit3rd@gmail.com (whit3rd)
Injection-Date: Thu, 23 Nov 2023 17:56:58 +0000
Content-Type: text/plain; charset="UTF-8"
Content-Transfer-Encoding: quoted-printable
X-Received-Bytes: 1826
 by: whit3rd - Thu, 23 Nov 2023 17:56 UTC

On Wednesday, November 22, 2023 at 3:05:15 PM UTC-8, pe...@easthope..ca wrote:
> Hi,
>
> The photo here is one of a pair of Eppindorf 5415D centrifuges.
> http://easthope.ca/Eppindorf5415D.jpg
>
> This morning both machines failed to operate. .... Fuses are OK.

Open 'em up and look for more fuses. There are modular
parts in many appliances, and the main fuse might be slower
or less sensitive than the fuse in one of the modules.
I've seen three fuses in series, and 'fusible resistors', reset-button
circuit breakers... safety consideration can get excessive.

1
server_pubkey.txt

rocksolid light 0.9.81
clearnet tor